114
Z31900_00_01
Error Codes
117.310.033 Security Eject failed. A sample was detected in the 
magazine but was expected to 
be in the magnet.
• Check sensors working.
• Adjust sample detection sensor 
threshold (Bruker service only).
117.310.034 Activate Release 
failed.
The sample release lever could 
not be activated to insert the 
sample into the magnet.
• Check if sample conveyor chain 
link release lever at this position is 
working correctly.
117.310.037
117.310.081
117.310.082
Insert failed. The sample could not be 
inserted into the magnet, e.g. 
was not detected down in the 
magnet.
• Check if sample gets stuck some-
where on its way down into the 
magnet; 
• Adjust the sample down detection 
sensor (Bruker service only).
117.310.039 Sample Detect at 
Magazine failed.
The sample was ejected from 
the magnet, but was not 
detected up in the magazine.
• Check if there was really a sample 
in the magnet;
• Adjust the sample down detection 
sensor threshold (Bruker service 
only);
• Clean the sensor detection sur-
faces; Adjust sensor detection 
threshold (Bruker service only).
117.310.067 No more measured 
samples available.
There are no more samples in 
the sample pool which are 
marked as „measured“.
117.310.111 Adapting cassette 
failed.
The cassette could not be 
adapted successfully.
• Try again.
117.310.112 Closing Clutch failed. The mechanical adaptation of 
the driving unit failed. The 
clutch could not be closed.
• Try again. Use another cassette.
117.310.114 Open Cassette 
Clamp failed.
The cassette clamp which fixes 
the cassette could not be 
opened successfully.
• Lift the front of the cassette a few 
millimeters and try again.
117.310.118 Close Guide failed The movable shell half of the 
guide did not reach its closed 
position.
• Check guide operation manually.
• Check if chain position is really 
correct; 
• Check if something (e.g. broken 
samples) prevents smooth guide 
motion.
117.310.122 Deactivate Release 
failed.
The sample release lever could 
not be deactivated.
• Check if sample conveyor chain 
link release lever at this position is 
working correctly.
103.310.145
117.310.123
Sample missing at 
Insert Position.
The requested insert position is 
empty.
• Check if there is really no sample.
• Clean sensor surfaces. 
• Adjust sample detection sensor 
threshold (Bruker service only).
117.410.011 Sensor Error: Cas-
sette: both sensors 
active!
The internal plausibility check of 
the sensors failed. The system 
stopped to avoid any damage 
to the equipment.
Contact Bruker for replacing the 
defective parts.