EasyManua.ls Logo

Chroma A620028 - Page 107

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
3-59
Figure 3-102
2. When the address of the SLAVE to be calibrated is confirmed, stay in parallel mode and
go to the CONFIG setting page, press then to enter into CALIBRATION
for password input as shown in Figure 3-100.
3. Select MASTER OR SLAVE via theRotary Knob” (
) and select the SLAVE to be
calibrated as shown in Figure 3-103. Press to enter into the calibration screen
as shown in Figure 3-104. For the calibration procedure for each item, see the
descriptions in sections 3.3.7.1 to 3.3.7.3.
Figure 3-103
Figure 3-104
1. A password is required for CALIBRATION. The password is “3636”.
2. A SLAVE model can only be calibrated via the MASTER. First
remove the output cable and then disconnect the current sharing
cable. Leave the ANALOG and SYSTEM BUS cables on the device.
3. The MASTER provides the panel output for the SLAVE. The load
and measurement cables for the actual calibration need to be
installed on the SLAVE.
4. The DC Power Supply should be calibrated every year. Contact
Chroma for further information.
7
ENTER
ENTER

Table of Contents

Related product manuals