EasyManua.ls Logo

Chroma A620028 - Page 109

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
3-61
4. In the Voltage calibration page, press “ to confirm.
5. The low voltage range calibration is done first. The instrument will set the output voltage
to 8.00V and the cursor stops at position [1] as shown in Figure 3-107. Enter the voltage
measured by the DVM and press “ ” to confirm.
6. Press “
” again to do the low voltage range calibration for the second point. The
instrument will set the output voltage to 115.00V and the cursor stops at position [2] as
shown in Figure 3-107. Enter the voltage measured by the DVM and press “ ” to
confirm.
[ V O L T A G E C A L I B R A T I O N ]
R E M O V E E X T E R N A L L O A D A N D P R E S S [ E N T E R ]
S E T T I N G O U T P U T V O L T A G E 8 . 0 0 V
A C T U A L
O U T P U T V O L T A G E = 7 . 9 9 9 8 _ V
S E T T I N G O U T P U T V O L T A G E 1 1 5 . 0 0 V
A C T U A L
O U T P U T V O L T A G E = 1 1 4 . 9 8 V
[ 1 ]
[ 2 ]
Figure 3-107
7. Press “
” to do the high voltage range calibration. The instrument will set the
output voltage to 150.00V and the cursor stops at position [3] as shown in Figure 3-108.
Enter the voltage measured by the DVM and press “ ” to confirm.
8. Press “
” again to do the high voltage range calibration for the second point. The
instrument will set the output voltage to 525.00V and the cursor stops at position [4] as
shown in Figure 3-108. Enter the voltage measured by the DVM and press “ to
confirm.
[ V O L T A G E C A L I B R A T I O N ]
R E M O V E E X T E R N A L L O A D A N D P R E S S [ E N T E R ]
S E T T I N G O U T P U T V O L T A G E 1 5 0 . 0 0 V
A C T U A L
O U T P U T V O L T A G E = 1 4 9 . 9 8 _ V
S E T T I N G O U T P U T V O L T A G E 5 2 5 . 0 0 V
A C T U A L
O U T P U T V O L T A G E = 5 2 4 . 9 8 V
[ 3 ]
[ 4 ]
Figure 3-108
9. The voltage calibration is now done. To save the calibration data, press to
display the confirmation page as shown in Figure 3-109. Press “ ” or turn the
“Rotary” (
) to set SAVE=YES and press “ ” to save it. If there is no need to save
it, press “
EXIT
” to return to the Calibration screen.
ENTER
ENTER
ENTER
ENTER
ENTER
SAVE
1
ENTER

Table of Contents