Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
5-38
4. PROGram:RUN
Description: Executes or stops the program.
Syntax: PROGram:RUN ON
PROGram:RUN OFF
Parameter: ON/1, OFF/0
Example: PROG:RUN ON
Query Syntax: PROGram:RUN?
Return Parameter: <NR1>
Query Example: PROGram:RUN?
Return Example: 1
5. PROGram:SAVE
Description: Saves the program.
Syntax: PROGram:SAVE
Parameter: None
Example: PROG:SAVE
6. PROGram:PULL
Description: Executes the PROGRAM TRIGGER input signal to enhance the
resistance control.
Syntax: PROGram:PULL <ARG>
Parameter: LOW|HIGH
Example: PROGram:PULL LOW
PROGram:PULL HIGH
Query Syntax: PROGram:PULL?
Return Parameter: LOW | HIGH
Query Example: PROGram:PULL?
Return Example: LOW
7. PROGram:SEQuence:SELected
Description: Sets the execution sequence of a program.
Syntax: PROGram:SEQuence:SELected <NR1>
Parameter: 1 to 10
Example: PROG:SEQ:SEL 3
Query Syntax: PROGram:SEQuence:SELected?
Return Parameter: <NR1>
Query Example: PROG:SEQ:SEL?
Return Example: 3
8. PROGram:SEQuence:TYPE
Description: Sets the sequence TYPE.
Syntax: PROGram:SEQuence:TYPE TRI
PROGram:SEQuence:TYPE AUTO
PROGram:SEQuence:TYPE MANUAL
Parameter: SKIP, AUTO, MANUAL
Example: PROG:SEQ:TYPE TRI
PROG:SEQ:TYPE AUTO
PROG:SEQ:TYPE MANUAL
Query Syntax: PROG:SEQ:TYPE?
Return Parameter: SKIP, AUTO, MANUAL
Query Example: PROG:SEQ:TYPE?
Return Example: 1