Remote Operation
5-39
9. PROGram:SEQuence:VOLTage
Description: Sets the voltage output sequence.
Syntax: PROGram:SEQuence:VOLTage <NRf+>
Example: PROG:SEQ:VOLT 40.5
Query Syntax: PROG:SEQ:VOLT?
Return Parameter: <NRf+>
Query Example: PROG:SEQ:VOLT?
Return Example: 4.050000e+01
10. PROGram:SEQuence:VOLTage:SLEW
Description: Sets the output voltage slew rate sequence.
Syntax: PROGram:SEQuence:VOLTage:SLEW <NR1>
Parameter: 0.01 to 10.00
Example: PROG:SEQ:VOLT:SLEW 1
Query Syntax: PROG:SEQ:VOLT:SLEW?
Return Parameter: <NR1>
Query Example: PROG:SEQ:VOLT:SLEW?
Return Example: 1.000000e+01
11. PROGram:SEQuence:CURRent
Description: Sets the output current sequence.
Syntax: PROGram:SEQuence:CURRent <NRf+>
Example: PROG:SEQ:CURR 40.5
Query Syntax: PROG:SEQ:CURR?
Return Parameter: <NRf+>
Query Example: PROG:SEQ:CURR?
Return Example: 4.500000e+01
12. PROGram:SEQuence:CURRent:SLEW
Description: Sets the output current slew rate sequence.
Syntax: PROGram:SEQuence:CURRent:SLEW <NRf1>
Example: PROG:SEQ:CURR:SLEW 10
Query Syntax: PROG:SEQ:CURR:SLEW?
Return Parameter: <NR1>
Query Example: PROG:SEQ:CURR:SLEW?
Return Example: 1.000000e+00
13. PROGram:SEQuence:CURRent:SLEWINF
Description: Sets the output current slewrate sequence to INF.
Syntax: PROGram:SEQuence:CURRent:SLEWINF ENABLE
PROGram:SEQuence:CURRent:SLEWINF DISABLE
Parameter: ENABLE/DISABLE
Example: PROGram:SEQuence:CURRent:SLEWINF ENABLE sets the
Slewrate to INF
PROGram:SEQuence:CURRent:SLEWINF DISABLE releases the
INF Slewrate and returns to 1A/ms
Query Syntax: PROGram:SEQuence:CURRent:SLEW?
Return Parameter: INF. Or <NRf+>[Unit Amp]
Query Example: PROGram:SEQuence:CURRent:SLEW? Returns the setting to
default.
Return Example: INF.
14. PROGram:SEQuence:TIME
Description: Sets the time duration of a sequence.