Alarms
86 Instructions for use Oxylog 3000 plus SW 1.n
Silencing acoustic alarm signals
Press the key (A).
The alarm indicator remains active and all
current acoustic alarm signals are silenced for
about 2 minutes. For alarms with a higher
priority, alarm silence is disabled, i.e. they are
signaled by an acoustic alarm signal.
After the 2 minutes have elapsed, all alarm
tones are reactivated.
To terminate the alarm silence before the 2 minutes
have elapsed:
Press the key (A) again.
In case of power supply failure
In the event of a power supply failure, ventilation,
volume measurement and alarms do not operate.
An acoustic alarm signal sounds to indicate the
power supply failure.
Spontaneous breathing can continue through the
emergency air inlet.
Immediately start ventilating the patient with an
independent manual ventilation device (manual
resuscitator) and maintain ventilation as
necessary using PEEP and/or increased
inspiratory oxygen concentration.
WARNING
Risk of alarms being unnoticed
Check the display regularly for alarm mes-
sages when the acoustic alarm signals are
silenced.
038
NOTE
To be notified of new acoustic alarm signals, the
2 minutes alarm tone silencing must be reset.