Proline Promass E 500 PROFIBUS PA Operation
Endress+Hauser 169
Parameter Prerequisite Description User interface Factory setting
S&W volume flow For the following order code:
"Application package", option
EJ "Petroleum"
The API referenced
correction option is selected
in Petroleum mode
parameter.
The software options
currently enabled are
displayed in the
Software option
overview parameter.
Displays the S&W volume flow
which is calculated from the
measured total volume flow
minus the net volume flow.
Dependency
The unit is taken from:
Volume flow unit parameter
Signed floating-point
number
S&W correction value For the following order code:
"Application package", option
EJ "Petroleum"
The External value option
or Current input 1…n
option is selected in the
S&W input mode
parameter.
The software options
currently enabled are
displayed in the
Software option
overview parameter.
Shows the correction value for
sediment and water.
Positive floating-
point number
Reference density alternative For the following order code:
"Application package", option
EJ "Petroleum"
In the Petroleum mode
parameter, the API
referenced correction
option is selected.
The software options
currently enabled are
displayed in the
Software option
overview parameter.
Displays the fluid density at
the alternative reference
temperature.
Dependency
The unit is taken from:
Reference density unit
parameter
Signed floating-point
number
GSV flow For the following order code:
"Application package", option
EJ "Petroleum"
The API referenced
correction option is selected
in Petroleum mode
parameter.
The software options
currently enabled are
displayed in the
Software option
overview parameter.
Displays the measured total
volume flow, corrected to the
reference temperature and the
reference pressure.
Dependency
The unit is taken from:
Corrected volume flow unit
parameter
Signed floating-point
number
GSV flow alternative For the following order code:
"Application package", option
EJ "Petroleum"
In the Petroleum mode
parameter, the API
referenced correction
option is selected.
The software options
currently enabled are
displayed in the
Software option
overview parameter.
Displays the measured total
volume flow, corrected to the
alternative reference
temperature and the
alternative reference pressure.
Dependency
The unit is taken from:
Corrected volume flow unit
parameter
Signed floating-point
number