16 Libraries FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 115
The input of a material or a medium is almost identical. Therefore, displays for a medium
will only be shown and described in case of differences.
Material properties
Select Special Funct.\Install Material or In-
stall Medium. Press ENTER.
This error message will be displayed if the coefficient
memory does not contain an area for user defined materi-
als/media.
Partition the coefficient memory accordingly (see section
16.1).
Select edit. Press ENTER.
Select a user defined material/medium. Press ENTER.
Change the designation of the material/medium.
The default name for a user defined material/medium is
USER MATERIAL N or USER MEDIUM N with N being an
integer.
Note! 95 ASCII characters (letters, capital letters, numbers, special char-
acters [!? " + - ( ) > < % * etc.]) are available for the designation of
materials/media.
A designation can have max. 16 characters. The input of text is de-
scribed in section 4.4.
Enter the sound speed of the material. Press ENTER.
For the sound speed of some materials annex C.1.
Enter the roughness of the material. Press ENTER.
For the typical roughness of some materials see annex
C.2.
Special Funct.
Install Material
USER Material
NOT FORMATTED !
Install Material
>EDIT< delete
USER Material
#01:--not used--
EDIT TEXT (↑↓← 
USER MATERIAL 1
c-Material
1590.0 m/s
Roughness
0.4 mm