88 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x 12 Displaying the Measured Values
12.5 Transducer Distance
The optimum transducer distance (here: 51.2 mm) is displayed in parentheses in the up-
per line, followed by the entered transducer distance (here: 50.8 mm).
The optimum transducer distance might change during the measurement (e.g. due to
temperature fluctuations).
A deviation from the optimum transducer distance (here: -0.4 mm) is compensated in-
ternally.
By pressing key during the measurement, it is possible
to scroll to the display of the transducer distance.
Note! Never change the transducer distance during the measurement!
L=(51.2) 50.8 mm
54.5 m3/h
DISP