22 Troubleshooting FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 163
22 Troubleshooting
If any problem appears which can not be solved with the help of this manual, contact our 
sales office and give a precise description of the problem. Specify the type, the serial 
number and the firmware version of the transmitter.
Calibration
FLUXUS is a very reliable instrument. It is manufactured under strict quality control, us-
ing modern production techniques. If installed as recommended in an appropriate loca-
tion, used cautiously and taken care of conscientiously, no troubles should appear. The 
transmitter has been calibrated at the factory and, usually, a re-calibration of the trans-
mitter will not be necessary. A re-calibration is recommended if
• the contact surface of the transducers shows visible wear or
• the transducers were used for a prolonged period of time at a high temperature (sever-
al months >130 °C for normal transducers or > 200 °C for high temperature transduc-
ers).
The transmitter has to be sent to FLEXIM for recalibration under reference conditions.
The display does not work at all or fails regularly
Check if the battery is inserted and charged. Connect the power supply. If the power sup-
ply is ok, the transducers or an internal component of the transmitter are defective. The 
transducers and the transmitter have to be sent to FLEXIM for repair.
The message SYSTEM ERROR is displayed
Press key BRK to return to the main menu.
If this message is displayed repeatedly, write down the number in the lower line. Track 
down the situations when the error is displayed. Contact FLEXIM.
The backlight of the display does not work, but all other functions are available.
The backlight is defective. This problem does not affect the other functions of the display. 
Send the transmitter to FLEXIM for repair.
Date and time are wrong, the measured values are deleted when the transmitter is 
switched off
The data backup battery has to be replaced. Send the transmitter to FLEXIM.