76 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x 11 Basic Measurement
11.2.6 Medium Pressure
The medium pressure is used for the interpolation of the sound speed and of the gas
compressibility factor.
11.3 Other Parameters
11.3.1 Transducer Parameters
If transducers are detected on a measuring channel, the parameter input finished. Press
ENTER. The main menu will be displayed.
If no or special transducers are connected, the transducer parameters have to be en-
tered.
11.4 Selection of the Channels
The channels on which will be measured can be activated individually.
Enter the medium pressure. Press ENTER.
This display will only be indicated if Special Funct.\SYSTEM
settings\Measuring\Gas-Measuring is activated
OR if Gas-Measuring is deactivated and Special
Funct.\SYSTEM settings\Dialogs/Menus\Fluid
pressure is activated.
Select Standard to use the standard transducer parame-
ters stored in the transmitter.
Select Special Version to enter the transducer param-
eters. The transducer parameters must be provided by the
transducer manufacturer.
Press ENTER.
Note! If standard transducer parameters are used, FLEXIM can not guar-
antee for the precision of the measured values. A measurement
might even be impossible.
If Special Version has been selected, enter the 6
transducer parameters specified by the manufacturer.
Press ENTER after each input.
Select program branch Measuring. Press ENTER.
If this error message is displayed, the parameters are not
complete. Enter the missing parameters in the program
branch Parameter.
Fluid pressure
60.00 bar
Transducer Type
Standard
Transd. Data 1
35.99
par>MEA<opt sf
Measuring
par>MEA<opt sf
NO DATA!