17.4 Settings of the Standard Conditions for the Gas Measurement
Select normal to display and transmit the profile corrected
flow values, uncorr. to display and transmit the flow val-
ues without flow profile correction. Press ENTER.
Fur further information see section 13.5.
A lower limit for the flow velocity can be entered (see sec-
tion 13.4).
An upper limit for the flow velocity can be entered (see
section 13.3).
Enter 0 (zero) to deactivate the flow velocity check.
Select the overflow behavior of the totalizers (see section
13.2.1).
Select on to keep the previous totalizer values after a re-
start of the measurement.
Select off to reset the totalizers to zero after a restart of
the measurement.
The activation of the turbulence mode can improve the sig-
nal quality if the flow is highly turbulent (e.g. in the vicinity
of an elbow or valve). An SNR value of min. 6 dB is re-
quired during the measurement.
Note! All changes will be stored at the end of the dialog.
Select Special Funct.\SYSTEM settings\Gas-
Measuring. Press ENTER.
This display will only be indicated if the gas measuring has
been activated in Special Funct.\SYSTEM set-
tings\Measuring.
Enter the pressure for the local standard conditions.
Enter the temperature for the local standard conditions.
Flow Velocity
>NORMAL< uncorr.
Cut-off Flow
absolut >SIGN<
Cut-off Flow
factory >USER<
Velocity limit
24.0 m/s
Quant. wrapping
off >ON<
Quantity recall
off >ON<
Turbulence mode
off >ON<
SYSTEM settings
Gas-Measuring
Normal pressure
1.01325 bar
Normal temper.
0.0 C