18 SuperUser Mode FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 129
18 SuperUser Mode
The SuperUser mode offers the possibility of an advanced analysis of the signal and the 
measured values as well as the definition of additional parameters adapted to the mea-
suring point, in order to achieve better measuring values or during experimental work. 
Features of the SuperUser mode are:
• Defaults will not be observed.
• There are no plausibility checks when parameters are being entered.
• There is no check whether the entered parameters are within the limits determined by 
the laws of physics and technical data.
• The cut-off flow is not active.
• A value for the number of sound paths must be entered.
• Some menu items that are not visible in the normal the normal mode are displayed.
18.1 Activation/Deactivation
Enter HotCode 071049 immediately after the transmitter has been switched on.
The SuperUser mode is deactivated by switching off the transmitter.
18.2 Transducer Parameters
In the SuperUser mode, the menu item Transducer Type will be displayed at the end 
of the parameter input, even if the transducers are detected by the transmitter.
Attention! The SuperUser mode is intended for experienced users with ad-
vanced application knowledge. The parameters can affect the nor-
mal measuring mode and lead to wrong measuring values or to a 
failure of the measurement when a new measuring point is set up.
It is displayed that the SuperUser mode is activated. Press 
ENTER. The main menu will be displayed.
Attention! Some of the defined parameters are still active after the deactivation 
of the SuperUser mode.
Press ENTER.
or
Select Special Version to enter the transducer param-
eters. Press ENTER.
SUPERUSER MODE
*IS ACTIVE NOW*
Transducer Type ↕
Q2E-314
Transducer Type ↕
Special Version