22 Troubleshooting FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 167
22.6 Problems with the Totalizers
The values of the totalizer are too high
See Special Function\SYSTEM settings\Measuring\Quantity recall. If
this menu item is activated, the values of the totalizer will be stored. The totalizer will con-
tinue with this value at the start of the next measurement.
The values of the totalizer are too low
One of the totalizers has reached the upper limit and has to be reset to zero manually.
The sum of the totalizers is not correct
See Special Function\SYSTEM settings\Measuring\Quant. wrapping. The
sum of both totalizers (throughput) transmitted via an output is not valid after the overflow
(wrapping) of one of the totalizers.
22.7 Data Transmission
The file with the transmitted measuring data contains meaningsless strings
The transmission parameters of the transmitter and the transmission program are not
identical. Adjust the transmission parameters of the transmitter (see section 14.2.4) and
of the program FluxData (see section 14.2.7) or of the terminal program.