21 Outputs FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 153
21.3 Activation of an Analog Output
21.3.1 Measuring Range of the Analog Outputs
After an analog output has been activated in the program branch Output Options, the 
measuring range of the source item must be entered.
Note! An output can only be activated in the program branch Output Op-
tions if it has been previously installed.
In the program branch Output Options, select the 
channel for which an output is to be activated. Press EN-
TER.
This display will not be indicated, if the transmitter has only 
one measuring channel.
Press ENTER until Current Loop is displayed. Select 
yes to activate the output. Press ENTER.
Select  sign if the sign of the measured values is to be 
considered for the output.
Select absolut if the sign is not to be considered.
Enter the lowest expected measured value. The unit of 
measurement of the source item will be displayed.
Zero-Scale Val. is the measured value that corre-
sponds to the lower limit of the output range as defined in 
section 21.1.1.
Enter the highest expected measured value.
Full-Scale Val. is the measured value tha corre-
sponds to the upper limit of the output range as defined in 
section 21.1.1.
example: output: current output
output range: 4…20 mA
Zero-Scale Val.: 0 m
3
/h
Full-Scale Val.: 300 m
3
/h
volumetric flow rate = 0 m
3
/h, corresponds to 4 mA
volumetric flow rate = 300 m
3
/h, corresponds to 20 mA
Output Options ↕
for Channel A:
Current Loop
I1: no >YES<
Meas.Values
>ABSOLUT< sign
Zero-Scale Val.
0.00 m3/h
Full-Scale Val.
300.00 m3/h