21 Outputs FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 149
21.1.1 Output Range
21.1.2 Error Output
In the following dialog, an error value can be defined which is to be output if the source
item can not be measured e.g. if there are solids in the medium.
When configuring an analog output, the output range will
be defined now. Select a list item or other range... to
enter the output range manually.
If other range... has been selected, enter the values
Output MIN and Output MAX. Press ENTER after each
input.
This error message will be displayed if the output range is
not min. 10 % of the max. output range. The next possible
value will be displayed. Repeat the input.
example: I
MAX
- I
MIN
2 mA for a 4…20 mA current output
Tab. 21.2: Error output
error value result
Minimum output of the lower limit of the output range
Hold last value output of the last measured value
Maximum output of the upper limit of the output range
Other value...
The value must be entered manually. It must be within the limits of the output.
I1 Output range ↕
4/20 mA
I1 Output MAX ↕
12.0 MINIMAL