21 Outputs
If the transmitter is equipped with outputs, they have to be installed and activated before 
they can be used:
• assign a measuring channel (source channel) to the output (if the transmitter has more 
than one measuring channel)
• assign the physical quantity (source item) to be transmitted to the output by the source 
channel, and the properties of the signal
• define the behavior of the output in case no valid measured values are available
• activate of the installed output in the program branch Output Options
21.1 Installation of an Output
All outputs are installed in Special Funct.\SYSTEM settings\Proc. outputs.
Note! The configuration of an output will be stored at the end of the dialog. 
If the dialog is quit by pressing key BRK, the changes will not be 
stored.
Select  Special Funct.\SYSTEM settings\Proc. 
outputs. Press ENTER.
Select the output to be installed. Press ENTER.
The scroll list contains all available outputs. A tick  after a 
list item indicates that this output has already been in-
stalled.
This display will be indicated if the output has not been in-
stalled yet. Select yes. Press ENTER.
If the output has already been installed, select no to recon-
figure it or yes to uninstall the output and to return to the 
previous menu item to select another output. Press EN-
TER.
Select in the scroll list the measuring channel to be as-
signed to the output as the source channel. Press ENTER.
This display will not be indicated, if the transmitter has only 
one measuring channel or only one measuring channel is 
active.
SYSTEM settings ↕
Proc. outputs
Install Output ↕
Current I1
I1 Source chan. ↕
Channel A: