B Technical Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 191
Dimensions
inputs
The inputs are galvanically isolated from the transmitter.
number max. 4
accessories input adapter (if number of inputs > 2)
temperature input
designation Pt100/Pt1000
connection 4-wire
range -150 ...+560 °C
resolution 0.01 K
accuracy ±0.01 % of reading ±0.03 K
current input
accuracy 0.1 % of reading ±10 A
passive input R
i
= 50 , P
i
< 0.3 W
- range -20...+20 mA
voltage input
range 0...1 V
accuracy 0.1 % of reading ±1 mV
internal resistance R
i
= 1 M
FLUXUS
in mm
FLUXUS G601
G601
U L T R A S O N I C F L O W M E T E R
1 3
2
0
4 65
7
8
N E X T
Q
O N
Q -
Q +
M U X
D I S P
Q
O F F
L I G H T
E N T E R
C
O N
C H A N N E L A
C H A N N E L B
9
D I S P
E N T E R
3 x O F F
M O D E
B R K
B A T T E R Y