220 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x B Technical Data
Pipe Damping Mat Type A: Quantity Depending on the Outer Pipe
Diameter
outer pipe diameter number of pieces
mm for 1 reflexion point for 2 reflexion points
100 4 7
200 7 13
300 10 19
500 16 32
1000 32 63
Pipe Damping Mat Type B: Length l Depending on Transducer Frequency
and Outer Pipe Diameter
outer pipe diameter transducer frequency
mm G, H K, M, P
300 12 m 6 m
500 32 m 16 m
1000 126 m 63 m