34 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x 6 Installation of FLUXUS G601
Storing the Battery
The battery remains in the transmitter. After storage, the transmitter can immediately be
operated with the battery.
• charge state: > 30 %
• storing temperature: 12...25 °C
Maintenance (Relearn Cycle)
The accuracy of the displayed value for the charge state of the battery is improved by ex-
ecuting a relearn cycle. The ambient temperature during a relearn cycle should be in the
range 12...30 °C.
Proceed as follows for a relearn cycle:
• Charge the battery completely. The LED "BATTERY" lights green when charging is fin-
ished.
• Discharge the battery completely: Remove the power supply unit from the transmitter.
To deactivate the automatic power off during discharging, start a measurement. Dis-
charging takes min. 14 h. The LED "BATTERY" will flash red afterwards.
Automatic Power Off
In the battery mode, the transmitter has an automatic power off. The transmitter will be
switched off if
• no measurement is being made and no key is pressed in 10 min or
• the battery is empty
Select Special Funct.\Battery status. Press
ENTER.
The charge state of the battery is displayed (here: 73
%).
The "?" and RELEARN indicate that the displayed
charge state is not reliable. A relearn cycle is recom-
mended.
This message will be displayed before the transmitter is
switched off automatically. A countdown with an acous-
tic signal will be started.
The countdown can be stopped by pressing any key.
If this message is displayed when the transmitter is
switched on, the transmitter has been switched off auto-
matically due to a too low charge state.
Special Funct. ↕
Battery status
■ ■ ■ ■ ■ ?73‰-
RELEARN! Cy: 24
■ LOW BATTERY
WHILE POWER OFF