46 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x 7 Installation of FLUXUS G608
The charge state of the battery can be displayed during the measurement (see section
10.3) and in the program branch Special Funct.:
This message will be displayed if the battery is almost empty:
Charging the Battery
• Connect the power supply unit to the transmitter (see Fig. 7.8).
• Switch on the transmitter.
The charging starts automatically. The LED "BATTERY" flashes green while charging.
The max. charging time is approx. 5 h.
During the charging process, the ambient temperature should be in the range 0...60 °C.
A measurement can be made during the charging. Charging will be stopped automatical-
ly when the battery is fully charged. The LED "BATTERY" will light green.
Select Special Funct.\Battery status. Press
ENTER.
The current charge state of the battery is displayed
(here: 30 %).
The minus sign "-" indicates that the transmitter is in
battery mode and is being discharged.
The number of cycles the battery has passed is dis-
played after Cy:.
A cycle corresponds to a charging and discharging pro-
cess. The life time of the battery can be derived by
means of this value.
If RELEARN is displayed in the lower line and a question
mark "?" is displayed in front of the current charge state,
a relearn cycle should be started (see section Mainte-
nance on the following page).
The capacity is sufficient for the display and storing of
the current parameter record. A measurement is not
possible anymore.
Attention! Observe the Safety Instructions for the Use in Explosive Atmo-
sphere (see document SIFLUXUS_608).
Special Funct. ↕
Battery status