Remote Operation
SCPI Commands and Syntax 5
5-33
Example:
1.0E1,1.141E1,0.0E0,0.0E0
Note:
P is identical to WATT.
S is identical to VA.
SOURce:PHASe<x>:POWer:FRYZe? [<cpd>{ P | S | Q }]
This command is a query only. The default (no parameter) version will return all of the
components from a calculation of the Phase’s Power according to Fryze in comma separated
format, in the order P, S, Q. The parameters select a specific component to return. The
returned numbers will be in standard scientific format.
Example:
1.0E1,1.141E1,0.0E0
Note:
P is identical to WATT.
S is identical to VA.
SOURce:PHASe<x>:POWer:KUSTers? [<cpd>{ P | S | QC | QCR | QL | QLR }]
This command is a query only. The default (no parameter) version will return all of the
components from a calculation of the Phase’s Power according to Kusters & Moore in
comma separated format, in the order P, S, Qc, Qcr, Ql, Qlr. The parameters select a
specific component to return. The returned numbers will be in standard scientific format.
Example:
1.0E1,1.414E1,0.314E0,0.1E0,0.207E0,0.207E0
Note:
P is identical to WATT.
S is identical to VA.
SOURce:PHASe<x>:POWer:SHEPherd? [<cpd>{ P | S | SR | SX | SD }]
This command is a query only. The default (no parameter) version will return all of the
components from a calculation of the Phase’s Power according to Shepherd & Zakikhani in
comma separated format, in the order P, S, Sr, Sx, Sd. The parameters select a specific
component to return. The returned numbers will be in standard scientific format.
Example:
1.0E1,1.414E1,1.314E0,0.1E0,0.0E0
Note:
P is identical to WATT.
S is identical to VA.
SOURce:PHASe<x>:POWer:SHARon? [<cpd>{ P | S | SQ | SC }]
This command is a query only. The default (no parameter) version will return all of the
components from a calculation of the Phase’s Power according to Sharon & Czarnecki in
comma separated format, in the order P, S, Sq, Sc. The parameters select a specific
component to return. The returned numbers will be in standard scientific format.