Remote Operation
Worked Examples 5
5-79
Setup frequency. SOUR:FREQ 60
Setup phase 1 (master) voltage range. SOUR:PHAS1:VOLT:RANG 0,180
Set amplitude (in absolute units). SOUR:PHAS1:VOLT:MHAR:HARM1 110,0
Set amplitudes and phases of harm 3. SOUR:PHAS1:VOLT:MHAR:HARM3 10,0
Set amplitudes and phases of harm 5. SOUR:PHAS1:VOLT:MHAR:HARM5 5,90
Enable main harmonics. SOUR:PHAS1:VOLT:MHAR:STAT ON
Enable voltage output (phase 1). SOUR:PHAS1:VOLT:STAT ON
Set output to on (all phases). OUTP:STAT ON
Example 6.
The harmonics in the previous example can be cleared before setting up new parameters so
that they do not interfere with any new setup. This can be a useful approach, if a full *RST
is not convenient.
Note: This will also clear the fundamental.
Clear all the harmonics. SOUR:PHAS1:VOLT:MHAR:CLE
Example 7.
This example shows how to setup a 110 VRMS 60 Hz voltage output from the voltage
terminals and a 1 A, 60 Hz current output from the current terminals. The current output
lags the voltage output by 90 °.
Reset all parameters to a known state. *RST
Ensure Output is off. OUTP:STAT OFF
Set voltage units to absolute. UNIT:MHAR:VOLT ABS
Set current units to absolute. UNIT:MHAR:CURR ABS
Setup frequency. SOUR:FREQ 60
Setup phase 1 (master) voltage range. SOUR:PHAS1:VOLT:RANG 0,180
Set Amplitude (in absolute units). SOUR:PHAS1:VOLT:MHAR:HARM1 110,0
Setup phase 1 (master) current range. SOUR:PHAS1:CURR:RANG 0.2,2
Set amplitude and phase. SOUR:PHAS1:CURR:MHAR:HARM1 1,-90
Enable voltage output. SOUR:PHAS1:VOLT:STAT ON
Enable current output. SOUR:PHAS1:CURR:STAT ON
Set output to on (all phases). OUTP:STAT ON