6100B/6105A
Users Manual
5-80
Example 8.
The previous example can be duplicated using a master unit to produce the voltage and an
auxiliary unit to produce the current.
Reset all parameters to a known state. *RST
Ensure output is off. OUTP:STAT OFF
Disable voltage output (master). SOUR:PHAS1:VOLT:STAT OFF
Disable current output (auxiliary). SOUR:PHAS2:CURR:STAT OFF
Setup Frequency. SOUR:FREQ 100
Setup phase 1 (master) voltage range. SOUR:PHAS1:VOLT:RANG 0,180
Set amplitude (in absolute units). SOUR:PHAS1:VOLT:MHAR:HARM1 110,0
Setup phase 2 (aux) current range. SOUR:PHAS2:CURR:RANG 0.2,2
Set amplitude and phase. SOUR:PHAS2:CURR:MHAR:HARM1 1,-90
Enable voltage output. SOUR:PHAS1:VOLT:STAT ON
Enable current output. SOUR:PHAS2:CURR:STAT ON
Set output to on (all phases). OUTP:STAT ON
Example 9.
Configure a master unit to output a pure 5V dc signal. Note that as dc is treated by the
6100B as the 0
th
harmonic of a fundamental frequency, that frequency must be defined even
when as in this case, the fundamental has zero amplitude.
Also note that previously enabled phenomena e.g., dip or interharmonic etc. would remain
enabled if the *RST was not commanded.
Reset all parameters to a known state. *RST
Use abs units for voltage harmonics. UNIT:MHAR:VOLT ABS
Setup Phase 1 (master) voltage range. SOUR:PHAS1:VOLT:RANG 1.1,16
Enable main harmonics (needed for DC). SOUR:PHAS1:VOLT:MHAR:STAT ON
RMS amplitude (to remove fundemetal). SOUR:PHAS1:VOLT:MHAR:AMPL 0
DC amplitude (will calculate new RMS). SOUR:PHAS1:VOLT:MHAR:HARM0 5,0
Setup the fundamental frequency. SOUR:FREQ 60
Enable voltage output on this phase. SOUR:PHAS1:VOLT:STAT ON
Set output to on (all phases). OUTP:STAT ON