6100B/6105A
Users Manual
7-10
ANDARD MASTER
VOLTAGE OUTPUT ON
CURRENT OUTPUT ON
1430V
Pk
MAX
30A Pk
15V Pk
MAX
46V Pk
MAX
VOLTAGE
CURRENT
OUTPUT
OUTPUT
SENSE
1V Pk
MAX
1V Pk
MAX
46V Pk
MAX
HI
LO
HI
LO
HI
LO
6105A ELECTRICAL POWER ST
STBY
DIRECT
MODE
SELECT
MENU
OPER
ESC
ENTER
0
1
4
7
.
2
5
8
EBKSP SHIFT
3
6
9
NEXT
CHAR
ALPHA
LOCK
_
/ ABC DEF
GHI JKL MNO
PQRS TUV WXYZ
3
4
5
6
ENERGY COUNTER
INPUTS
1
2
0 - 28 V MAX
Power
(4 Wire)
Sense
IHIH
OLOL
I
(2 Wire)
Input
Terminals
Front
Rear
Guard
Open
To LO
1A/250V
500Vpk
MAX
Guard
200Vpk
MAX
200Vpk
MAX
1000Vpk
MAX
DCV
DCI
ACV
ACDCV OHMF
Range
Auto
Menu Scroll
Display / Window
Hold Test Reset Address
ACDCI
C
E
Menu
N
Error
S
Clear
?
Recall
L
R
T
PER
OHM
ACI FREQ
FUNCTION / RANGE
MENU
NUMERIC / USER
Local
8
7
9
E
5
4 6
-
2
1 3
Back
Space
.
0
'
Enter
NPLC
Auto
Cal
Offset
Comp
Auto
Zero
N Rdga/
Trig
Trig
Store
State
Recall
State
On
Off
NULL STAT MATH OFF
MEAN SDEV UPPER
LF OFF
SYNC ACV
SCALE LOWER
HP3458A/HFL
All Term 1000pk Max
Transducer
6100BDigitizing DMM
Trigger
gdw56.eps
Figure 7-3. Phase Measurement Connections
Figure 7-3 shows how the 6100B, a digitizing DMM and transducer are connected. Note
that voltage dividers and shunts are used to scale the input to the DMM for optimum
performance. The same DMM is used to calibrate voltage and current.