6100B/6105A
Users Manual
7-18
Range
(Amps)
Fundamental
Frequency
(Hz)
Harmonic
Number
Frequency
(Hz)
Point
Output
Level
Setting
(Amps)
Output Phase
Angle Setting
(Degrees)
0 0
DC
offset
0.000 0
1 57 Gain 18.000 0
8 456 Gain 18.000 0
15 855 Gain 18.000 0
57
27 1539 Gain 16.000 0
48 3024 Gain 16.000 0
69 4347 Gain 10.000 0
80 5040 Gain 10.000 0
21
63
95 5985 Gain 10.000 0
1 57 Gain 45.000 0
6 342 Gain 45.000 0
15 855 Gain 45.000 0
57
48 3024 Gain 20.000 0
50
63 95 5985 Gain 20.000 0
1 57 Gain 64.000 0
6 342 Gain 64.000 0
15 855 Gain 64.000 0
57
48 3024 Gain 20.000 0
80
63 95 5985 Gain 20.000 0
Table 7-10. Current Adjustment Points for 80 A Option (if fitted)
Range
(Amps)
Frequenc
y
(Hz)
Harmonic
Number
Setting
6100B
Contribution
Measurement
Std. Deviation
Combined
Verification
Tolerance (High)
Combined
Verification
Tolerance (Low)
Result
8 – 0.80
57 1
Amplitude 64 A ±8 mA
Phase 0° ±0.0005°
2961 47 Amplitude 64 A ±11 mA
Phase 0° ±0.002°