Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
10
Document no. M1144953-001
− Replace the Ref. filter assembly.
− Replace the filters in the pneumatic unit (1 or 2 pcs).
− Check the D-fend O-rings and replace them, if necessary.
− Replace the D-fend and sampling line.
− Clean or replace the fan filter.
NOTE: Use only approved sampling lines to ensure proper functioning.
Use a 2 m/7 ft. sampling line with Compact Airway Modules E-/ M-COVX and E-/ M-CAiOVX.
3. The manufacturer recommendation is to replace the CO
2
absorber every 4 years. Replace
the absorber, if necessary.
NOTE: All Compact Airway Modules do not contain a CO
2
absorber.
"
Connect the Compact Airway Module to a module slot with a long gas interface cable (the
grounding plates of the cable should be removed for the procedure).
2.1.3 Single-width Airway Module, E-miniC/ M-miniC
1. Detach the module box.
2. Check that all cables and tubes are connected properly and that there are no loose
objects inside the module.
3. Install the PM Kit:
− Replace the special tube (Nafion™).
− Replace the Zero valve air filter every three years.
− Check the D-fend body connector O-rings and replace them, if necessary.
− Replace the Mini D-fend and the sampling line.
NOTE: Use only approved sampling lines to ensure proper functioning.
4. Connect the module to a module slot with a long gas interface cable (the grounding
plates of the cable should be removed for the procedure).
NOTE: Noisy sampling pump might indicate possible problems with motor bearing. Replace the
noisy sampling pump by new one if needed.
"
2.1.4 Tonometry Module, E-TONO/ M-TONO
1. Detach the module box.
Check that all cables and tubes are connected properly and that there are no loose
objects inside the module.
2. Replace the Special tube (Nafion 2 pcs).
Reattach the module box and check that the latch moves properly. Plug the module back
into the module frame.
"