Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
28
Document no. M1144951-004
option software, the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the
same monitor, but not onto any other monitor.
1. Make sure that the monitor is switched to standby.
2. Open the cover for card drive slots (on the left side of the monitor). Insert the option
software card into one of the card drive slots and press the software card firmly in
position.
3. Switch the monitor on.
4. Wait until the normal screen appears and check that a message 'Field Upgrade OK' is
displayed as a note of successful upgrade.
5. Remove the option software card.
6. Switch the monitor to standby.
7. Detach the cover for card drive slots and install the antenna cover.
8. Insert the wireless LAN antenna card with adapter into the foremost card drive slot.
9. Switch the monitor on.
10. Enter the service menu and make sure that the option was downloaded successfully
(Monitor Setup - Install/Service - Service - Frame - Options - "WLAN" or "All").
11. Enter the WLAN service menu (Monitor Setup - Install/Service (16-4-34)- Service
(26-23-8)- Frame - Network - WLAN - WLAN Config).
Enter appropriate Network ID, WEP Algorithm, Key ID and Encryption Key using the
corresponding menu selections.
For selecting allowed communication channels, enter the appropriate Channel Mask
using the corresponding menu selection.
See Appendix C. Channel Mask Selections.
Save the monitor’s wireless LAN configuration by selecting Save Configuration - Save.
NOTE: The settings must match up with the ones that are defined in the Access Point
configuration. Refer to the related site configuration documentation for correct settings, if
necessary.
12. Check all installed options on the monitor software device plate and attach the Virtual ID
sticker under the device plate at the monitor rear panel.
13. Test the monitor on Datex-Ohmeda Network.
14. Fill out all necessary documentation regarding the upgrade.
3.3.12 Installing the S/5 Recorder Upgrade, U-CMREC1
NOTE: Only S/5 Compact Anesthesia Monitor rev. 02 or later and S/5 Compact Critical Care
Monitor rev. 02 or later can be upgraded with this S/5 Recorder Upgrade kit.
1. Turn the monitor off.
2. Disconnect the power cord.
3. Push the Service Reset button for at least 5 seconds until the service reset indicator LED
turns off (the fan should turn off).
4. Remove the plate that covers the recorder installation place in the side panel.
5. Open the recorder (by pressing the button on the right upper corner at the front of the
recorder).
6. Insert the recorder from outside to the side panel (opening button upwards).