Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
40
Document no. M1144955-001
4.7.7 NIBP Watchdog
How to test Adult watchdog time
Select Test ADULT and press the ComWheel. Watchdog interval starts counting up seconds
and keeps on counting as long as the +15 Vdirty is on. The time should be 120 to 140 seconds.
How to test Infant watchdog time
Select Test INFANT and press the ComWheel. Watchdog interval starts counting up seconds
and keeps on counting as long as the +15 Vdirty is on. The time should be 60 to 70 seconds.