Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
58
Document no. M1144951-004
Others -Resp Setup -Size -1.0
- Resp Rate Source – AUTO
- Measurement – ON
- Detection Limit – AUTO
5.3.2 Display
1. Check that the picture on the screen is  correct. 
"
5.3.3 Keyboard(s)
1. Tests with all the connected keyboards:
− Press the 
Monitor Setup key. Turn the ComWheel in both directions and check that 
the cursor in the menu moves correspondingly. Select Normal Screen and check 
that the menu disappears from the screen.
Tests with the Command Board:
− Press the 
Help key. Turn the ComWheel in both directions and check that the cursor 
in the menu moves correspondingly. Select Normal Screen and check that the 
menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
Tests with the ARK Keyboard:
− Enter the Keyboard service menu
− Check functioning of the ComWheel.
− Press all keys, except 
Modify and Print. Check that each key produces a sound 
from the loudspeaker, or the Message count value in the service menu increases.
− Press the 
Modify and Print keys and check that the corresponding menus appear 
on the monitor screen.
Tests with the Remote Controller:
− Enter the Keyboard service menu.
− Check the function of the ComWheel.
− Press all keys. Check that each key produces a sound from the loudspeaker, or the 
Message count value in the service menu increases.
"
5.3.4 Frame unit
1. Check that the frame fan is running.
"
5.3.5 Display unit
1. Check that the clock on the screen shows correct time. Readjust the time and date, if 
necessary.
"