Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
22
Document no. M1144951-004
disappears, or in STBY mode the Stby LED stops blinking (may take up to 10 hours if the
batteries are fully discharged).
WARNING The power cord may only be connected to a three-wire, grounded, hospital
grade receptacle.
3.3.3 Connecting to Network
Make sure that the Datex-Ohmeda Network Option, N-CMNET or S/5 MemCard and Network
Option, N-CMMEM is installed.
NOTE: With L-CANE03(A) and L-CICU03(A) or newer software, the N-CMNET option is
automatically installed.
Use the Monitor-Network cable to connect the monitor to the network.
1. Make sure that the power is switched off.
2. Connect the Identification Plug and one RJ-45 connector to the connectors X4 and X5.
3. Connect the other RJ-45 connector to the corresponding connector on the wallbox.
4. Switch on the power. Confirm that the network symbol and 'Connected to Network'
message are displayed on the upper part of the screen.
3.3.4 Connecting to Wireless Network
Make sure that the S/5 Wireless LAN Option, N-CMW and the antenna cover are installed. Use
the wireless LAN antenna card to connect the monitor to the network.
NOTE: The N-CMW option requires also that either N-CMNET or N-CMMEM option is installed.
1. Make sure that the power is switched off.
2. Connect the wireless LAN antenna card into the foremost card drive slot.
3. Switch on the power. Confirm that the wireless LAN network symbol and the wireless LAN
signal strength symbol are displayed on the upper part of the screen.