Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4
Document no. M1144953-001
1.2.1 Hemodynamic patient simulators
The following tables present the patient simulators’ compatibility with each hemodynamic
module, and the accessories needed:
Table 1 Patient simulators’ compatibility with each hemodynamic module
Table 2 Adapter cables for hemodynamic patient simulators
Module Parameter
Patient simulator
M1010831 MedSim 874027
E-PSM(P),E-PRESTN ECG Cable included Multilink ECG acc. Not compatible
T 402015-004 402015-004 and
M1010832
InvBP Cable included M1010858
M-PRESTN ECG Ok Multilink ECG acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
M-NESTPR ECG Not compatible 300 series acc. Ok
12 lead ECG Not compatible 300 series acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
Patient simulator
Hemodynamic patient simulator Dual temperature adapter cable 402015-004
Hemodynamic patient simulator Dual Inv.BP adapter cable 2005772-001
Medsim Temperature adapter cable M1010832
NIBP
ECG
SpO
T1
T2
P1
P2
2
Auto
On/Off
Start
Cancel
Zero
P1
Zero
P2