EasyManua.ls Logo

GE Datex-Ohmeda S/5 Series - Figure 18 S;5 Compact Monitor General Troubleshooting Flowchart

GE Datex-Ohmeda S/5 Series
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General troubleshooting
65
Document no. M1144951-004
6 General troubleshooting
Figure 18 S/5 Compact Monitor general troubleshooting flowchart
Monitor not
functioning
Power cord
connected?
Connect power
cord
Monitor can be
started now?
Batteries were
empty
On/Standby LED
is ON or blinks?
PMB failure.
Replace PMB.
Monitor can be
started now?
Display Unit
failure.
No
Yes
Yes
Yes
No
Possible AC/DC
unit failure.
Replace AC/DC
unit.
No
No
Frame Unit fan is
running?
Yes
No
AC/DC Unit
failure.
Yes
CM1_trbl_Flowchart.vsd

Table of Contents

Related product manuals