Interfacing
51
Document no. M1144951-004
E-PSM / E-PSMP. The ECG signal from the modules is channel 1 (ECG1). The channels 2 and 3
(ECG2, ECG3) are not transmitted. Make sure that the signal in channel 1 (ECG1) is good enough
(extensive QRS for IABP synchronization). For further information, please contact your
authorized GE Healthcare distributor.
Pressure out (X8 pin 4)
− P1 from Hemodynamic Module
NOTE: With
monitor software version 03 or earlier, the Invasive pressure output signal is 1
V/100 mmHg, originally ranging from 0 to 300 mmHg and with a delay of approximately 25ms.
The signal requires an input impedance of 100k.
4.7.3 S/5 Pressure Temp Module, E-PT or M-PT, output signals
The signal output connector on the Pressure Temp Module, E-PT or M-PT, can be used to
interface some models of IABPs to the S/5 Compact Anesthesia Monitor and S/5 Compact
Critical Care Monitor. The pin assignments are illustrated in Table 13. Please contact your local
distributor for more information.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
Table 13 Signal output connector pin assignments
Output signals
ECG out, ECG1 1V / 1mV
Pressure out, P3 1V / 100mmHg
ECG out, ECG1 1V / 1mV
Pin Signal
1
2
3
4
ECG out, ECG1
Pressure out, P3
ECG out, ECG1
Ground