Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2
Document no. M1144958-001
1Specifications
Dimensions 335 x 205 x 340 mm /13.2 x 8.1 x 13.4in
(W x D x H)
Weight 8.5 kg (19.4 lbs) (w/o modules)
Electrical Requirements
Power supply
Rated voltage range and frequencies 100 to 240 VAC 50/60 Hz
(limits 90-264 VAC)
Stability ±10 % of nominal voltage
Power consumption 140 VA (max)
Fuses 2x2A, T/SLOW, 250 V
Grounding Hospital grade
Interruptability Data memory and alarm settings are saved up to 15 minutes
after monitor standby
Battery
Battery type NiMH, 3.5Ah
Battery charging time 10 hours to full capacity
Battery operation up to 1.5 hours typical on full charge at 23 °C / 73 °F
Recharging when connected to mains power
Battery operaton alarm
early warning ‘Battery low’ note
final warning ‘Battery low’ alarm
Battery state indicator Capacity bar displayed on the screen
Environmental Requirements
Operating temperature 10 - 35 °C / 50 - 95 °F
Storage temperature -10 - 50 °C / 14 - 122 °F
Atmospheric pressure 500 - 800 mmHg (660 - 1060 mbar, 66 - 106kPa)
Humidity 10 - 90 % non-condensing
In-built recorder
Power consumption 3 W
Principle Thermal array
Print resolution
Vertical 8 dots/mm (200 dots/inch)
Horizontal 32 dots/mm (800 dots/inch) at a speed of 25 mm/s and slower
Paper width 50 mm, printing width 48 mm
Traces Selectable 1, 2, or 3 traces
Print speed 1, 6.25, 12.5, 25 mm/s