Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
50
Document no. M1144955-001
HW Module hardware status. See Table 3 Module hardware status for a detailed description of the
message.
Table 3 Module hardware status
Module key Module keyboard status. See Table 4 Module keyboard status for a detailed description of the
message.
Table 4 Module keyboard status
Error Statuses (HEX):
General error Module general error status. See Table 5 Module general error status for a detailed description
of the message.
Table 5 Module general error status
bit 0 Lamp ON
bit 1 Stepper motor ON
bit 2 Zeroing valve ON
bit 3 Pump valve ON
bit 4-15 Not used
bit 0 Button 1 (start/stop) pressed
bit 1 Button 2 (Lab.Data) pressed
bit 2-15 Not used
Bit Description What to do
bit 0-1 Not used. -
bit 2 Zeroing error. Zero gas signal out of limits or
pressure or temperature signal varying during
measurement.
CO
2
sensor faulty. Send the
module to the factory for
repair.
bit 3 CO
2
measurement error. CO
2
signal out of limits or
pressure or temperature signal varying during
measurement.
CO
2
sensor faulty. Send the
module to the factory for
repair.
bit 4 Measurement error. Sensor temperature,
measurement pressure or absorbance is out of
limits.
CO
2
sensor faulty. Send the
module to the factory for
repair.
bit 5 Calibration error. Failure in calibration. Recalibrate.
bit 6 Pneumatic error. See pneumatic error status
bits for details.
bit 7 Lamp pulsing error. CO
2
sensor faulty. Send the
module to the factory for
repair.
bit 8-15 Not used. -