Display Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
3
Document no. M1144956-003
Encryption Wired Equivalent Privacy (WEP) 40 and 128 bit
WLAN adapter PC card type with antenna
F-CM(C)1..05 and newer CF card with adapter
1.6 Recorder option
Power consumption 3 W
Printer type Thermal array
Print resolution Vertical 8 dots/mm (200 dots/inch)
Horizontal 32 dots/mm (800 dots/inch)
at a speed of 25 mm/s and slower
Paper width 50 mm, printing width 48 mm
Traces Selectable 1, 2, or 3 traces
Print speed 1, 6.25, 12.5, 25 mm/s
1.7 Digital outputs of Display Unit
Nurse call Low State: 0-0.8V; High State: 2.8-5V
Generated on red and yellow alarms
Defibrillation Synchronization State 0: 0-0.8V; state 1: 2.8-5V
Delay (max): 35 ms from R peak
Signal active time 10 ms
Serial Output Asynchronous
Serial data interface, uses RS-232 standard
Baud rate max 115.2 kbps
Parity: None; Data bits: 8; Stop bit: 1