Sensor Service data is changed to the sensor view.
Entropy refers to the respective module key. Is
highlighted when the Entropy key is pressed on the
module. (Opens the Entropy menu.)
Check Sensor is highlighted when the Check Sensor
key is pressed on the module. Activates the sensor
check.
Cable indicates whether the Entropy sensor cable is
connected to the module or not.
Sensor indicates whether the Entropy sensor is
connected to the cable or not.
Sensor S/N module reads the manufacturing serial
number from the sensor chip.
Sensor Lot module reads the manufacturing lot
number from the sensor chip.
Lead 1/2/3 shows the lead on/off status of all sensor
electrodes.
Imped. 1/2/3 shows the respective impedance value of each
sensor electrode of the last impedance check.
Imp. Meas. shows On if the Automatic check (sensor impedance
measurement) is on in the module.
Imp. Check shows OK if the last impedance check has been
successful.