Document no. M1144955-001
Last Errors: The section shows the last monitor
errors and the time of their occurence.
Error counters: The section contains counters for
each different (detected) monitor error. The time of
occurrence of the last error is shown beside each
counter.
Last log reset: The date and time of the last Error/
Event log reset.
Possible errors:
Fast cold start indicates the number of erroneous
cold starts with power off time less than 20 seconds.
The reason can be either a failing lead acid battery
or a software problem that was solved by the
hardware watchdog circuitry located in the
monitor’s power supply unit.
GSP watch-dog timeout indicates an erroneous
restart controlled by the main CPU board and
caused by the display controller board. The restart is
listed as a fast cold start.
UPI watch-dog timeout indicates an erroneous restart controlled by the main CPU board and caused by the
UPI section. The restart is listed as a fast cold start.
Last events: The section shows the last events and
the time of their occurrence.
Event counters: The section contains counters for
each different (detected) event. The time of
occurrence of the last event is shown beside each
counter.
Last log reset: The date and time of the last Error/
Event log reset.
Possible events:
Cold start is a start-up with power off time more
than 15 minutes. The trend memory is cleared and
monitoring starts with the user default settings.
Warm start is a start-up with power off time less
than 15 minutes. The trend information and
possible temporary settings are still available.