Document no. M1144955-001
Start Test: A selection to start the automatic system
test sequence. The procedure is guided online.
Stop Test: A selection to stop the automatic system
test sequence.
Record Data: A selection to print out the shown
service data to the device defined in the ...Service -
Record Data menu.
TONO
Cathpres; pressure measured from the catheter line
[mbar].
Ambpres; ambient pressure in the room [mbar].
Pumppres; pressure measured from the pumping
unit [mbar].
System test results; detailed information of the test
results of different system parts. The meaning of the
messages:
- “OK”, the part successfully passed the test
- “FAIL”, the part failed in the test
- “N/A”, the test could not be carried out properly.
See Table 8 Description of system test for a detailed description of the tested parts.