4.7.2 NIBP Calibration
How to Calibrate
Calibration selection is available only when protection is OFF.
NOTE: Both channels B1 and B2 must be calibrated simultaneously.
NOTE: The module must be in the frame during the whole procedure.
NIBP calibration can be performed in the NIBP Service menu as follows:
1. If Protection is ON, change it to OFF by pressing the NIBP module buttons
Auto ON/OFF
and
Start Cancel simultaneously for 3 seconds, which enables the Protection selection.
Then press the buttons again for 3 seconds to enable Calibration.
NOTE: When the buttons have been pressed, the NIBP field shows an error message
‘Calibration switch on!’.
NOTE: For M-NE(12)STPR series modules: If Protection is ON, change it to OFF by first turning
the toggle switch to the right at the bottom of the module, which enables the Protection
selection. Then turn the toggle switch to the left to enable Calibration. When the switch is at
the right, the NIBP field shows an error message ‘Calibration switch on!’
NOTE: When calibration is enabled, a message ‘Calibration not protected’ appears.
2. For proper zeroing to take place, remove the hose from the front panel connector. Select
Calibration and push the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in the
NIBP message field. After this, a pressure bar will appear beside the menu and the text
‘Calibrating’ will appear in NIBP digit field.