Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
42
Document no. M1144951-004
4.2.5 Functional check
There are two ways to check the function of the Device Interfacing Solution:
1. Press the 
Monitor Setup key.
2. Select Interfacing and open the Status Page menu. The status page shows you the 
current communication status of the interfacing module (1 - 10). 
NOTE: The status message ‘Connected’ appears on the Status Page after you have connected 
the external device to the interfacing module and turned it on. Note also that the monitor and 
the interfacing module must be operational. 
• Check the LED indicators on the interfacing module (the green LED indicates physical 
connections, the yellow LED software selections)
:
4.2.6 Selecting the parameter data source
Select the external device via Monitor Setup - Interfacing menu:
• Select the desired parameter, for example Gas e s .
• Select the desired source by name, for example  Aestiva.
NOTE: The name of the device is visible on the list only if the device is correctly connected.
Detailed information about interfacing module related mountings, connections and settings is 
included in the installation guides that are delivered with the interfacing modules.
 GREEN  YELLOW  INDICATION
 lit              dark      
Physical connections between the monitor, interfacing 
module   and external device are in order and the device 
has been selected in the menu.
 dark
 
  lit            
There is something wrong with the physical connections 
between the monitor, interfacing module and external 
device. The external device has not been selected in the 
menu.
  lit                lit             
Physical connections between the monitor, interfacing 
module and external device are in order but the external 
device has not been selected in the menu.
 dark          dark         
The interfacing module is not connected to the monitor.