Table 15 Adapter cables for hemodynamic patient simulators
5.2 Visual inspection
Make sure that the monitor is switched to standby.
Disconnect the mains power cord from the monitor.
If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable
from the monitor. Remove any memory cards, if installed.
1. Check all units visually
Check that all parts are intact and that the cables and screws are connected and
tightened properly. Especially check the following parts:
− the display unit locking screw is in a locking position.
− sampling line is connected to the Airway Module, if installed.
Check that modules go in smoothly and lock up properly in all module slots.
M-PRESTN ECG Ok Multilink ECG acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
M-NESTPR ECG Not compatible 300 series acc. Ok
12 lead ECG Not compatible 300 series acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
Patient simulator Adapter cables for simulators
Hemodynamic patient simulator - Dual temperature adapter cable 402015-004
Hemodynamic patient simulator - Dual Inv.BP adapter cable 2005772-001
Medsim - Temperature adapter cable M1010832
Medsim - Inv.BP adapter cable M1010858
Lionheart & MPS450 - Temperature adapter cable M1010846
Lionheart & MPS450 - Inv.BP adapter cable M1010862
Module Parameter
Patient simulator
M1010831 MedSim 874027
NIBP
ECG
SpO
T1
T2
P1
P2
2
Auto
On/Off
Start
Cancel
Zero
P1
Zero
P2