Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
36
Document no. M1144951-004
Table 4 Transference of parameters, Datex-Ohmeda monitors
Table 5 Transference of parameters, external monitors
4.1.1 Connection to external Datex-Ohmeda monitors
Use the INT-External Device Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT /
M-INT. Tighten the finger screws.
Device Waveforms
(analog)
Numerics Alarms
Cardiocap E-INT / M-INT -> CO
2
Et&Fi Airway gases,
Respiration rate,
SpO
2
, Pulse rate
None
Capnomac
Capnomac II
E-INT / M-INT -> CO
2
,
Pleth
Et&Fi Airway gases,
Respiration rate
None
Capnomac Ultima E-INT / M-INT -> CO
2
Et&Fi Airway gases,
Respiration rate,
Spirometry, SpO
2
,
Pulse rate, Pleth
amplitude
CO
2
, O
2
, Anesthesia
agent, Respiration
rate, Apnea,
Occlusion, SpO
2
,
Pulse rate
Normocap 200
Normocap 200 OXY
E-INT / M-INT -> CO
2
Et&Fi Airway gases,
Respiration rate
None
Oscar
Oscar II
Oscar OXY
E-INT / M-INT -> CO
2
,
Pleth
Et&Fi Airway gases,
Respiration rate,
SpO
2
, Pulse rate,
Pleth amplitude
None
Satlite
Satlite II
Satlite Plus
E-INT / M-INT -> CO
2
SpO
2
, Pulse rate,
Pleth amplitude
None
Satlite Trans None SpO
2
, Pulse rate,
Pleth amplitude
None
Device Waveforms Numerics Alarms
Critikon
Dinamap 1846SX
None NIBP None
Abbott
Oximetrix 3
None SvO
2
/SaO
2
, CO None
Baxter
Explorer
None C.O., SvO
2,
REF,
Tblood
None
Baxter
Vigilance
None C.O., SvO
2,
C.C.O.,
Tblood
None
Nellcor
N-100
N-200
N-1000
Pleth (analog) SpO
2
, Pulse rate None