Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
16
Document no. M1144958-001
Frame
Power Supply
Battery
Check that the shown Battery Voltage value is within 10 – 14V.
"
18. Monitor temperature
Check that the shown Monit Temp value is within 20 - 40.
"
19. Battery operation
Disconnect the power cord (without switching the monitor to standby).
Check that the monitoring continues normally. The battery indicator should appear on
the upper right hand corner of the screen.
"
20. Mains power recognition
Reconnect the power cord.
Check that the LCD display screen brightness increases back to normal and the
monitoring continues without problems. The battery indicator should be replaced by the
text CHARGE.
"
3.2.6 Recorder Unit
21. Recorder messages
Open the recorder unit cover.
Check that the message ‘Recorder: Cover open’ appears on the screen, then close the
cover.
"
22. Recording
Press the key
Record Wave and check that the module starts recording the selected
waveforms. Press the key
Stop on the module to stop recording.
NOTE: InvBP waveforms are not recorded without a patient simulator.
NOTE: If no recording appears, check that the paper roll is installed in a correct way. Only one
side of the paper is printable.
"
23. Paper speed
Press the key
Record Wave again and this time wait until the recording stops. Check
that the length of the recorded waveform scale is 18.7 cm (1.5 cm).
Change the paper speed setting to 1 mm/s: