Display Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
41
Document no. M1144956-003
The communication is set in the monitor’s Network service menu.
Network software S-CNET99 -> DRI Level = 1999
Network software S-CNET01 -> DRI Level = 2001
Network software S-CNET02 -> DRI Level = 2001 or 2002 (WLAN)
Network software L-NET03 -> DRI Level = 2003
Network software L-NET05 -> DRI Level = 2005
NOTE: The DRI Level 1999 is not selectable in all monitor software versions.
NOTE: If the DRI level is changed, the monitor will restart automatically.
Fill out all necessary documentation regarding the new monitor software.
NOTE: The license agreement of monitor software needs to be in accordance with the monitor
software serial number. Make sure you archive the license agreement in a secure location.
NOTE: The first start-up after software loading takes considerably longer.
See also the troubleshooting chart for the software card, if necessary.
3.3.16 Performing Factory Reset
NOTE: The factory reset is necessary after replacing monitor software.
NOTE: The factory reset will restore all customized defaults, including language selection, to
factory defaults.
1. Press the
Monitor Setup key.
2. Select Install/Service and password (16-4-34).
3. Select Service and password (26-23-8).
4. Select Set/ Test and Factory Reset.
5. The monitor will perform an automatic restart. After the restart is completed, restart the
monitor also manually.
3.4 Adjustments and calibrations
It is not necessary to perform calibrations or adjustments on S/5 Compact Monitors.