EasyManua.ls Logo

GE Datex-Ohmeda S/5 Series - General

GE Datex-Ohmeda S/5 Series
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Frame Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
17
Document no. M1144958-001
Record/Print
Record Waveforms
Paper Speed --> 1 mm/s
Press the key
Record Wave and wait until the recording stops. Check that the length of
the scale is now 3.0 cm (± 0.5 cm).
"
24. Quality of recording
Check that the quality of the recordings is acceptable.
"
3.2.7 General
25. Service reset switch
Check the service reset switch.
Switch the monitor to standby and disconnect the power cord. Press the service reset
switch for at least five seconds. Check that the LED next to the switch turns off, and also
the fan turns off.
Reconnect the power cord and turn the monitor back on. Check that all trends have now
been erased.
"
26. Battery capacity
Typical lifetime of the Compact Monitor Ni-MH batteries is approximately 4 years.
Test the batteries by performing the following:
detach all modules
keep the monitor connected to the mains power until the batteries are fully charged
install a hemodynamic multiparameter module and disconnect the power cord
notice the time the monitor stays on with batteries
Replace the batteries, if the time was less than 45 minutes.
"
27. Battery charging
Reconnect the power cord, switch the monitor on and enter the Battery service menu:
Monitor Setup
Install/Service (Password 16-4-34)
Service (Password 26-23-8)
Frame
Power Supply
Battery
Check that the monitor starts charging one of the batteries and all data is reasonable.
"

Table of Contents

Related product manuals