Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
18
Document no. M1144958-001
28. Service Log reset
Enter the Service Log in the service menu:
Monitor Setup
Install/Service (Password 16-4-34)
Service (Password 26-23-8)
Service Log
Clear the content of the Service Log by selecting Reset Log in the menu.
"
29. Electrical safety check
Perform an electrical safety check and a leakage current test.
Check that the monitor functions normally after the performed electrical safety check.
"
30. Final cleaning
Switch the monitor to standby, disconnect the power cord and perform final cleaning.
After cleaning, reconnect the power cord for recharging the batteries.
"
Fill in all necessary documents.