17
Document no. M1144953-001
Change baseline impedance on the simulator and check that appropriate RESP waveform
and RR values are shown again within 30 seconds.
Turn the simulator off. Check that the ‘Asystole’ and ‘Apnea’ messages are displayed.
"
Temperature measurement
8. Enter the ESTP: STP service menu.
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the STP board memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all show OK.
9. Check the temperature measurement calibration using temperature test plugs.
NOTE: Make sure that the protection for temperature calibration is set on, if calibration is
needed.
"
Invasive blood pressure measurement
10. Check the function of the module front panel membrane keys.
11. Check the InvBP channels with a patient simulator.
Zero the InvBP channels, then check that the values and waveforms correspond to the
simulator settings.
"
SpO2 measurement
12. Check that the message ‘No probe’ is shown, when no SpO
2
sensor is connected.
Connect an SpO
2
finger probe to the module. Check that the message ‘Probe off’ is shown,
when the probe is not connected to a finger.
13. Attach the SpO
2
probe to your finger. Check that a reading of 95-99 and a pleth
waveform appear on the screen.
"
Non invasive blood pressure measurement
14. Enter the NIBP module service menu.
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the NIBP board memories have passed the
internal memory test, i.e. RAM, ROM and EEPROM all show OK.
15. Check the function of the front panel membrane keys.
16. Check the NIBP tubing system for leakages by performing Calibrations - Active leak test.
17. Perform NIBP calibration by selecting Calibration.
18. Enter the Pneumatics service menu.
Check the NIBP watchdog timer activation pressure with a pressure manometer.
19. For Modules M-ESTPR and M-NE(12)STPR: Check the NIBP watchdog timer by performing
Watchdog – Test Adult and Test Infant.
20. Check the safety valve by performing Safety Valve – Adult and Infant.