Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
20
Document no. M1144953-001
2.2.13 Pressure/Pressure Temp Modules, E-P, E-PT, M-P, M-PT
1. Enter the P/PT module service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the PT board memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all show OK.
"
Invasive blood pressure measurement
2. Check the function of the front panel Zero P3 membrane key.
3. Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
"
Modules w/ temperature measurement
4. Check the temperature measurement calibration using temperature test plugs.
NOTE: Make sure that the protection for temperature calibration is set on.
"
2.2.14 Dual pressure Module, E-PP, M-PP
1. Enter the PP module service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the Dual Pressure module, E-PP or M-PP,
memories have passed the internal memory test, i.e. RAM, ROM and EEPROM all show OK.
2. Check the function of the front panel
Zero P5 and Zero P6 membrane keys.
3. Check the InvBP channels with a patient simulator.
Zero the InvBP channels. Then check that the values and waveforms correspond to the
simulator settings.
"
2.2.15 Cardiac Output Modules, E-COP, E-COPSv, M-COP, M-COPSv
1. Enter the COP module service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the module’s memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all show OK.
"
Invasive blood pressure measurement
2. Check the function of the front panel membrane Zero P4 key.
3. Check the InvBP channel with a patient simulator.